AERIUS System Configurations
Aerius Automated Infrared Imaging System
- Includes the Aerius Instrument,
computer with LCD monitor, Aerius Application Software, and Aerius
Accessory Pack which includes a membrane carrier.
Aerius
Automated Infrared Imaging System—With Plate Stacker
- Includes
the Aerius Instrument, Bio-Stack Microplate Stacker, Bio-Stack
Plate Stacker install kit (includes stand, alignment tools and cables),
computer with LCD monitor, Aerius Application Software, and Aerius Accessory
Pack which includes a membrane carrier.
Aerius Specifications
Laser/Microscope
-
Aeruis Specifications
Laser/Microscope
- Laser Lifetime: 60,000 hours typical
- 700 Channel Laser Source: Solid-state diode laser at 685 nm
- 800 Channel Laser Source: Solid-state diode laser at 785 nm
- Detectors: Silicon avalanche photodiodes
- Scanning Speed: 6 minutes per microplate with 100 µm resolution
- Focusing Range: 2.25 µm (suitable for most microtiter plates with transparent, flat-bottom wells.)
Operation Specifications
- Operating Conditions: 15-35º C and dew point no greater than 20º C
- Power Requirements: Automatic voltage selection at 90-250 VAC and 47-63 Hz; 1.1 Amps at 120 V; 200 watts maximum
- Dimensions (instrument only): 36 h x 45 w 57 d cm (14 x 18 x 22 inches)
- Dimensions (instrument with stacker): 85 h x 110 w x 45 d cm (33 x 44 x 18 inches)
- Data Storage Capacity: > 25 GB
- Network Connection: Cat. 5 RJ-45, 10 Base-T/100 Base-TX
- Bar Code Reader symbology: Code 39 preferred. Compatiable with Code 128 and other symbologies. Contact LI-COR for more information