NUCLEIC ACID GEL DOCUMENTATION

Application Overview

Sensitivity is high and background fluorescence is low when using SYTO® 60 red fluorescent nucleic acid stain to visualize DNA gels on the LI-COR Odyssey® Infrared Imagers. The detection sensitivity and lower limit of detection for SYTO 60 with any of the Odyssey Imagers are better than those of ethidium bromide detected with either a Polaroid camera or a CCD imaging system1.

If you need to image ethidium bromide, SYBR® Safe DNA stains, or other DNA stains in your nucleic acid gels in the 600 nm region, please see the Odyssey Fc information on Nucleic Acid Gel Documentation.

dna staining figure 1

Figure 1.


Comparison of detection limits between ethidium bromide (Polaroid of CCD camera)
and SYTO 60 (Odyssey Infrared Imager).

DNA Fragment (bp) LOD (pg) LOD (fmol)
100 375 5.7
500 95 0.29
1500 44 0.044

Table 1. SYTO 60: Limit of Detection (LOD) with Odyssey Imaging

 

DNA Fragment (bp) LOD (pg) LOD (fmol)
100 3000 45
500 1500 4.5
1500 700 0.7

Table 2. Ethidium Bromide: Limit of Detection (LOD) with Polaroid Imaging

 

DNA Fragment (bp) LOD (pg) LOD (fmol)
100 750 11.4
500 375 1.14
1500 175 0.18

Table 3. Ethidium Bromide: Limit of Detection (LOD) with CCD Imaging


1. Syto® 60 Staining of Nucleic Acids on Gels LI-COR Biosciences (2010)

Technical Resources

Syto® 60 Staining of Nucleic Acids on Gels LI-COR Biosciences (2010)

 

Visit us on You TubeLike us on Facebook!Follow us on TwitterRed the blogger blog

Biotechnology
4647 Superior St Lincoln, NE 68504
Toll-Free: 800-645-4267
Email: biohelp@licor.com
Environmental
4421 Superior Street Lincoln, NE 68504
Toll-Free: 800-447-3576
Email: envsales@licor.com
© 2011 LI-COR Biosciences