If Comparing Film and Digital Imagers, Expose Blot on Digital Imager First.

If you are trying to compare how the same chemiluminescent Western blot looks when imaged on a digital imager (like the C-DiGit® Blot Scanner) with how it will look when imaged on film, it’s important to know that you should expose the blot to film BEFORE imaging on a digital imager.

Why does this matter? Digital imaging requires capturing the most photons being generated, which is typically immediately after a 5-minute chemiluminescent substrate incubation. Time may be more of an issue with some substrates. For more information on how film and digial imaging compare, read Western Blot Analysis: Comparison of film and the C-DiGit Blot Scanner.

In Table 1 below, performance differences of a Western blot detected with SuperSignal® West Pico when the same blot is imaged over time. Blot was incubated 5 min in substrate before imaging on the C-DiGit Blot Scanner. Images are normalized to the LUT of the optimal blot.

Table 1 Optimal Blot Unsatisfactory Blot Unsatisfactory Blot
Images Optimal Blot with SuperSignal West Pico Unsatisfactory Blot with West Pico Unsatisfactory Blot with West Femto
Conditions: Immediately after incubation with SuperSignal West Pico 26 min after incubation 51 min after incubation
Imaging Time Immediately after incubation with SuperSignal® West Pico 26 min after incubation 51 min after incubation
Scan Setting High High High
Performance LOD – 625 ng, Signal – 338 LOD – 625 ng, Signal – 114 LOD – 625 ng, Signal – 32.2

In Table 2, Performance differences of a Western Blot detected with SuperSignal West Dura1 when the same blot is imaged over time. Blot was incubated 5 min in substrate before imaging on the C-DiGit Blot Scanner. Images are normalized to the LUT of the optimal blot.

Table 2 Optimal Blot Satisfactory Blot Satisfactory Blot
Images Optimal with West Dura Satisfactory with West Dura Satisfactory with West Dura
Conditions:
Imaging Time Immediately after incubation with SuperSignal West Dura 24 min after incubation 48 min after incubation
Scan Setting High High High
Performance LOD – 156 ng, Signal – 12,300 LOD – 156 ng, Signal – 10,400 LOD – 156 ng, Signal – 9,090

In Table 3, Performance differences of a Western Blot detected with SuperSignal West Femto when the same blot is imaged over time. Blot was incubated 5 min in substrate before imaging on the C-DiGit Blot Scanner. Images are linked to the LUT of the optimal blot.

Table 3 Optimal Blot Satisfactory Blot Satisfactory Blot
Images Optimal Blot with West Femto Satisfactory Blot with West Femto Satisfactory Blot with West Femto
Conditions:
Imaging Time Immediately after incubation with SuperSignal West Femto 24 min after incubation 48 min after incubation
Scan Setting High High High
Performance LOD – 156 ng, Signal – 11,500 LOD – 156 ng, Signal – 8,120 LOD – 156 ng, Signal – 6,860

1Comparable to WesternSure® PREMIUM Chemiluminescent Substrate

Related posts:

Troubleshooting Chemiluminescent Western Blots: Possible Cause 4 for Weak Signals – Blot Processing

Sometimes life in the lab gets crazy, right? You are finishing a Western blot and you realize that you are supposed to be at an important lecture across campus in 10 min!! Or, your spouse calls to say that one of the kids needs to be picked up as soon as possible. Yikes! The challenge is that blots should be processed and detected on the same day. And, the secondary antibody should be incubated the day of imaging and fresh substrate added just before imaging. Is it that important to your results? Yes, it is and just to prove it, we did a few experiments.

In Table 1, we studied performance differences when the same blot is imaged immediately after processing vs. stored overnight dry and then imaged. In Table 2, we looked at performance differences when the same blot is imaged immediately after processing vs. stored overnight wet and then imaged. Blots in both tables were all imaged on the C-DiGit® Blot Scanner. (And, all images are normalized to the Lookup Tables (LUT) of the respective optimal blot.)

For both experiments, you can see that saving the blot to image the next day is not a very good choice. This is because the secondary antibody and/or the chemiluminescent Western blot substrate is not stable enough for acceptable photon emission when digitally images after the day it is applied.

Table 1 Optimal Blot Unsatisfactory Blot Unsatisfactory Blot
Images Optimal Chemiluminescent Western Blot Unsatisfactory Chemiluminescent Western Blot Unsatisfactory Chemiluminescent Western Blot
Conditions:
Substrate SuperSignal® West Dura1 SuperSignal West Dura1 SuperSignal West Dura1
Processing Time Same Day Next Day Next Day
Detection Process HRP secondary incubated, washed, and substrate added immediately before imaging. HRP secondary incubated, washed, and substrate added day before imaging. HRP secondary incubated, washed, and substrate added day before imaging, then re-incubated with HRP secondary and substrate added immediately before imaging.
Storage Conditions Blot stored overnight dry, at room temperature Blot stored overnight dry, at room temperature
Performance LOD – 640 ng LOD – None detected LOD – 1.25 μg
Table 2 Optimal Blot Unsatisfactory Blot Unsatisfactory Blot
Images Optimal Chemiluminescent Western Blot Unsatisfactory Optimal Chemiluminescent Western Blot Unsatisfactory Optimal Chemiluminescent Western Blot
Conditions:
Substrate SuperSignal® West Dura1 SuperSignal West Dura1 SuperSignal West Dura1
Process Time Same day Next day Next day
Detection Process HRP secondary incubated, washed, and substrate added immediately before imaging. HRP secondary incubated, washed, and substrate added day before imaging. HRP secondary incubated, washed, and substrate added day before imaging, then re-incubated with HRP secondary and substrate added immediately before imaging.
Storage Conditions Blot stored overnight wet in PBS, at room temperature Blot stored overnight wet in PBS, at room temperature
Performance LOD – 640 ng LOD – None detected LOD – 1.25 μg

1SuperSignal West Dura results are comparable to those obtained with WesternSure® PREMIUM Chemiluminescent Substrate.

For more hints and tips, stay tuned to future blog posts. And if you would like to try some FREE Western Blot Analysis Software, download Image Studio™ Lite today!

Related posts:

WesternSure® Chemiluminescent Western Blotting Reagents from LI-COR®

WesternSure Chemiluminescent Western Blotting ReagentsDetect your Western Blots with Confidence! Use NEW! WesternSure® Chemiluminescent Reagents from LI-COR!

Now, in addition the great imaging systems for chemiluminescent Western blots, LI-COR Biosciences offers chemiluminescent Western blotting substrates and HRP-conjugated secondary antibodies for use in performing your BEST chemiluminescent Western blots ever. And, the WesternSure Pen is used to annotate visible protein ladders prior to chemiluminescent Western blot detection.

WesternSure chemiluminescent Western blotting reagents offer the best performance available when compared to other competitive products on the market. WesternSure PREMIUM Chemiluminescent Substrate is a highly sensitive enhanced substrate for detecting horseradish peroxidase (HRP) on immunoblots.

WesternSure HRP-conjugated secondary antibodies (Goat anti-Mouse and Goat anti-Rabbit) are compatible with a variety of chemiluminescent substrates and are optimized for use with WesternSure PREMIUM chemiluminescent substrates.

Happy Blotting!