Possible Cause 10 for Weak Chemiluminescent Western Blot Signals: Diluting Substrates

westernsure-premium-926-95000Okay, I know, research budget money is tight and you want to make your reagents stretch as far as possible, but it really not a good idea to dilute your chemiluminescent Western blotting substrate.

Why? It’s because the rate of reaction is determined by the ratio of enzyme to substrate. Diluting substrates will dramatically impact the overall generation of light. Then, you will have to repeat the experiment, and you end up using more substrate anyway!

Optimal Blot Unsatisfactory Blot
Images Optimal Western Blot - Substrate Not Diluted Unsatisfactory Chemiluminescent Western Blot - Substrate Diluted
Conditions:
Substrate SuperSignal® West Dura1 SuperSignal® West Dura1
Substrate NOT diluted. Substrate diluted 1:1 (in water)
Performance LOD – 1.25 µg LOD – 2.5 µg

1Comparable to WesternSure® PREMIUM Chemiluminescent Substrate

So don’t skimp – use the substrate full strength the first time to ensure that you are seeing all of your protein bands. Or you might just have to repeat the experiment (and that will just cost you more time and money. . .)!

Here are the other nine possible causes of weak chemiluminescent Western blot signals: