4300 Specifications
Laser/Microscope
Laser Lifetime: 40000 - 60000 hours typical.
700 Channel Laser Source: Solid-state diode laser at 685 nm.
800 Channel Laser Source: Solid-state diode laser at 785 nm.
Detectors: Two silicon avalanche photodiodes.
Scanning Speed: Five selectable scan speeds to optimize image quality.
Electrophoresis Power Supply
Maximum Output: 5000V, 40mA, and 120W.
Operation Specifications
Operating Conditions: 15 - 35C and dew point not greater than 19 C.
Power Requirements: Automatic voltage selection at 90-250 VAC and 47-63 Hz. 3 amps at 120volts; 380 watts maximum.
Dimensions
4300S: 59H x 51W x 47D cm (23" x 20"x 18"). 102 cm (40") wide with door fully open.
4300L: 84H x 51W x 47D cm (33" x 20"x 18"). 102 cm (40") wide with door fully open.
Weight
4300S: 29 kg (64 lb).
4300L: 36.3 kg (80 lb).
Internal Server
Data Storage Capacity: 80 GB for data storage and server software.
Network Protocol: TCP/IP.
Network Connection: RJ-45, 10/100 BaseT.
Security: Password protected access.
Gel Electrophoresis Apparatus
Glass Height |
Separation Distance |
Resolution (bp) |
Run Time (h) |
Applications |
66 cm* |
56 cm |
1000-1200 |
|
Sequencing |
41 cm |
31 cm |
800 |
|
Sequencing |
25 cm |
15 cm |
500 |
|
Genotyping and Sequencing |
* Available on 4300L only.