Sequencing BAC DNA With Near-Infrared
Fluorescent Non-Nucleotide TerminatorsStephen C. Roemer, Vince R. Boveia, Craig M. Johnson and D. Michael Olive
LI-COR Inc., Biotechnology, 4647 Superior Street, Lincoln, NE 68504
IRD800-AcyATP
BAC Sequencing Protocol
Note: Avoid vortexing the samples
DNA samples should be stored at 4°C prior to sequencing.
Sequencing mixture:
DNA template: 1.5-2.0 µg
Thermo Sequenase®: 8 units
LI-COR AcyNTP mixtures: 2 µl/rx
Water: to 27 µlThermal Cycling
IRDye® infrared dye AcycloTerminators
- 95 °C for 3 minutes
- 95 °C for seconds
- 50 °C for seconds
- 72 °C for 2 minutes
- Repeat 2 to 4 for 60 cycles total
- 72 °C for 5 minutes
- 4 °C soak
Labeled Primers
- 95 °C for 3 minutes
- 95 °C for 30 seconds
- 50 °C for 30 seconds
- 70 °C for 2 minutes
- Repeat 2 to 4 for 60 cycles total
- 4 °C soak