Sequencing BAC DNA With Near-Infrared
Fluorescent Non-Nucleotide Terminators

Stephen C. Roemer, Vince R. Boveia, Craig M. Johnson and D. Michael Olive
LI-COR Inc., Biotechnology, 4647 Superior Street, Lincoln, NE 68504


Sequencing Results

 

 

Comparison of the performance of IRD800-labeled acyNTPs and IRD800-labeled ddNTP for DNA sequencing. The lane-to-lane band intensities for all four IRD800-labeled acyNTPs were equivalent to those obtained with IRD800-labeled ddNTPs throughout the sequence ladder.

image


Comparison of read length and accuracy to 1000 bases for two primer walking projects
using either dye-primer or dye-AcycloTerminator sequencing.

Dye-AcycloTerminator
Dye-Primer
Custom Primer Read Length
Accuracy (%)
Read Length
Accuracy (%)
Template p4CG9
T3
1157
99.1
1067
99.4
T7
1090
99.4
1099
99.8
967T7F
1126
98.9
1163
100.0
218T7R-1
1192
98.4
1120
99.9
914T3F
1059
99.4
1129
99.8
284T3R
1093
99.4
1120
100.0
1872T7F
1090
99.8
1174
100.0
1874T3F
1162
99.5
1063
99.8
310T7R
1020
98.4
1154
99.6
221T3R
1120
99.5
1138
99.9
2742T3F
1118
97.6
1039
99.9
251T3R
1178
99.2
1140
99.9
2844T7F
1083
98.8
1085
99.8
218T7R-2
1010
98.2
1083
99.6
3624T3F
1177
99.4
1147
99.4
211T3R
1066
99.6
1224
99.9
3701T7F
1121
99.4
1112
99.6
228T7R
1036
98.0
1053
99.4
Template pUC19
M13F
1005
98.9
1092
99.5
508F
1073
99.9
1198
99.8
928F
1062
98.9
1164
99.7
1243F
1093
99.1
1108
100.0
1585F
1040
99.2
1184
99.6
1991F
1040
99.4
1141
99.5
2286F
1037
99.4
1152
99.6
2482F
1116
99.7
1067
98.1
MEAN
1090
99.1
1123
99.7

 

x
A A A
line
800 | 1024 | max
Biotechnology
4647 Superior St
Lincoln, NE 68504
Toll-Free: 800-645-4267
Email: biohelp@licor.com
Environmental
4421 Superior Street
Lincoln, NE 68504
Toll-Free: 800-447-3576
Email: envsales@licor.com