Specifications
Laser/Microscope
- Laser Lifetime: 40000 - 60000 hours typical
- 700 Channel Laser Source:Solid-state diode laser at 685 nm
- 800 Channel Laser Source:Solid-state diode laser at 785 nm
- Detectors:Two silicon avalanche photodiodes
- Scanning Speed: Five selectable scan speeds to optimize image quality
Operation Specifications
- Operating Conditions:15-35° C and dew point no greater than 19° C
- Power Requirements: Automatic voltage selection at 90-250 VAC and 47-63 Hz. 3 amps at 120volts; 380 watts maximum
Size and Weight
-
Dimensions (instrument only):4300S: 59H x 51W x 47D cm (23" x 20"x 18"). 102 cm (40") wide with door fully open
4300L: 84H x 51W x 47D cm (33" x 20"x 18"). 102 cm (40") wide with door fully open -
Weight:4300S: 29 kg (64 lb)
36.3 kg (80 lb)

| Gel Electrophoresis Apparatus | ||||
|---|---|---|---|---|
| Glass Height | Separation Distance | Resolution (bp) | Run Time (h) | Applications |
| 66 cm* | 56 cm | 1000-1200 | 10 | Sequencing |
| 41 cm | 31 cm | 800 | 6 | Sequencing |
| 25 cm | 15 cm | 500 | 1.5 - 3 | Genotyping and Sequencing |