Specifications

Laser/Microscope

  • Laser Lifetime: 40,000 working hours typical
  • 700 Channel Laser Source:Solid-state diode laser at 685 nm
  • 800 Channel Laser Source: Solid-state diode laser at 785 nm
  • Detectors: Silicon avalanche photodiodes
  • Scanning Speed:6 minutes with 100 µm resolution
  • Resolution:20 - 500 µm
  • Focusing Range:0 - 3.95 mm (suitable for focusing through the bottom of compatible microplates with transparent, flat-bottom wells.)

Operation Specifications

  • Operating Conditions:15-35° C and dew point no greater than 20° C
  • Power Requirements: Automatic voltage selection at 90-250 VAC and 47-63 Hz; 1.1 Amps at 120 V; 200 W maximum
  • Data Storage Capacity: Internal hard drive (files also stored on computer)

Size and Weight

  • Dimensions (instrument only):36 h x 45 w 57 d cm (14 x 18 x 22 inches)
  • Weight:24 kg (53 lbs)
OdysseySa Specification