Environmental
|
Environmental
Resolution: 1 mm2 (1 mm x 1 mm scanning area).
Accuracy: Within ± 2% for samples >50 cm2.
Display Capacity: Area: 9,999,999.99 cm2, Length: 99,999,999.9 cm,
Width: 12.8 cm.
Display: 2 line x 16 character LCD.
Keyboard: Sealed, 24 key tactile response with audio feedback.
Real Time Clock: Year, month, day, hour, minute, second. Accuracy: ± 3
minutes per month (0 to 55 °C).
Internal Memory: More than 2.1MB of non-volitale FLASH. Measurement data is maintained even in absence of power.
Memory Capacity: From 65,000 to 125,000 individual leaf area measurements.
Communications: USB 1.1 Compliant, Full Speed, Bus rate @ 12 Mbps; RS-232 DTE, N,8,1,
Baud rate set @ 38400 bps, Hardware Flow Control.
Sample Dimensions: Width: 127 mm maximum, 1 mm minimum. Thickness:
8 mm maximum. Length: 1 meter maximum.
Scanning Speed: Up to 1 m/s. Length encoding cord drawing speed need not be constant.
Error message indicates if encoding cord is drawn too fast.
Power Requirement: Rechargeable 6V lead-acid battery; or 115/230VAC,
50/60 Hz, 20W maximum.
Battery Capacity: Typically 12-15 hours of continuous operation.
Recharging Time: 5-8 hours, instrument off. Charging circuitry is built
in.
Battery Voltage Sensor: Low battery warning occurs approximately 1 hour before automatically shutting off the instrument. Over-voltage warning occurs if battery becomes over-charged.
Scanning Head: LED Light Source: 6.4 mm from outer edge. Size: 30.5 cm overall length (12.0"). Weight: 680 g (1.5 lb.)
Operating Temperature: 0 to 55° C, 0 to 95% RH (noncondensing).
Storage Temperature: -20 to 55° C, 0 to 95% RH (noncondensing). Long term storage at high temperatures will shorten life of internal battery.
Scanning Head: LED Light Source: 6.4 mm from outer edge. Size: 30.5
cm overall length (12.0"). Weight: 680 g (1.5 lb.).
Readout-Control Unit: Size: 19.8 L x 15.5 W x 10.1
cm D (7.8" x
6.1" x 4.0"). Weight: 2.0 kg (4.4 lb.).
Resolution: 1 mm2 (1 mm x 1 mm scanning area).
Accuracy: Within 1% for samples >10 cm2.
Sample Dimensions: Width: 127 mm maximum, 1 mm minimum. Thickness:
7.5 mm maximum. Length: Unlimited.
Conveyor Belt Speed: 6.3 cm/s at 60 Hz; 5.4 cm/s at 50 Hz.
Transparent Belts: Rugged clear vinyl.
Power Requirement: 115/230 VAC, 50 to 60 Hz, 20 W maximum.
Operating Temperature: 0 to 55 ° C, 0 to 95% RH (noncondensing).
Storage Temperature: -20 to 55 ° C, 0 to 95% RH (noncondensing).
Size: 27.9 H x 27.9 W x 38.1 cm L (11.0" x 11.0" x 15.0").
Weight: 7.7 kg (17.0 lb.)
Specifications subject to change without notice