Environmental
|
Environmental
Resolution: 1 mm2 or 0.1 mm2 (adjustable)
Scanning Area: 1 mm2 Resolution: 1 mm x 1 mm; 0.1 mm2 Resolution: 0.300
mm W x 0.333 mm L
Display Capacity: 1 mm2 Resolution: 999,999.99 cm2; 0.1 mm2 Resolution:
99,999.999 cm2
Display: Full 8-digit LED
Width: 25.4 cm max; 1.5 to 3.0 mm min
Thickness: Up to 2 cm, user-expandable to 2.5 cm
Length: Unlimited
Conveyor Belt Speed: 8.0 cm/s at 60 Hz; 6.7 cm/s at 50 Hz
Light Source: 15 W fluorescent tube
Transparent Belts: Rugged clear vinyl
Power Requirements: 108-126/216-252 VAC, 48 to 66 Hz, 100 W max
Operating Temperature: +15 to + 55 °C
Storage Temperature: -20 to + 65 °C
Size: 25.0 H x 60.0 W x 73.0 L cm (9.8” x 23.6” x 28.7”)
Weight: 43 Kg (95 lb)
Accuracy
Sample Area |
||||
| Resolution | 10 cm2 | 3 cm2 | 1 cm2 | 0.3 cm2 |
| 1 mm2 | ± 2.0% | ± 3.0% | ± 6.0% | ± 10.0% |
| 0.1 mm2 | ± 1.0% | ± 1.5% | ± 3.0% | ± 5.0% |
Combined accuracy and precision to 99% confidence with correct calibration on verifiable shapes. Better accuracy can be achieved by calibrating the LI-3100C and/or placing the leaf on the middle section of the belt. Use the 0.1 mm2 resolution for conifers, roots and other similar objects and expect the accuracy to be about 5% less than normal leaves.
Specifications subject to change without notice