LI-820 Specifications
CO2 Specifications
Measurement Range: 0-1000 ppm, 0-2000 ppm with 14 cm bench; 0-5000 ppm, 0-20000 ppm with 5 cm bench
Accuracy: < 2.5% of reading with 14 cm bench; 4% of reading with 5 cm bench
Calibration Drift
1Zero Drift: < 0.15 ppm / °C
2Span Drift at 370 ppm: < 0.03% / °C
3Total Drift at 370 ppm: <0.4 ppm / °C
RMS Noise at 370 ppm with 1 sec Signal Filtering: < 1 ppm
1 Zero drift is the change with temperature at 0 concentration
2 Span drift is the change after re-zeroing following a temperature change
3 Total drift is the change with temperature without re-zeroing or re-spanning
Measurement Principle: Non-Dispersive Infrared
Traceability: Traceable gases to WMO standards from 0-3000 ppm. Traceable gases to EPA protocol gases from 3000 to 20000 ppm
Pressure Compensation Range: 15 kPa-115 kPa
Maximum Gas Flow Rate: 1 liter/minute
| Output Signals: | Two Analog Voltage (0-2.5 V or 0-5 V) and Two Current (4-20 mA) |
| Digital: TTL (0-5 V) or Open Collector | |
DAC Resolution: 14-bits across user-specified range
Source Life: 18000 hours
| Power Requirements: | Input Voltage 12-30 VDC |
| 1.2A @ 12V (14 W) maximum during warm-up with heaters on | |
| 0.3 A @ 12 V (3.6 W) average after warm-up with heaters on | |
Supply Operating Range: 12-30 VDC
Operating Temperature Range: -20 to 45 °C
Relative Humidity Range: 0 to 95% RH, Non-Condensing
Dimensions: 8.75" x 6" x 3" (22.23 x 15.25 x 7.62 cm)
Weight: 2.2 lbs (1 kg)