Infrared Gas Analyzer
The analyzer in the LI-8100 is an absolute, non-dispersive, infrared (NDIR) gas analyzer with a single path,
dual wavelength detection system. The optical bench provides a CO
2 measurement range of 0-3000 µmol/mol and
an H
2O measurement range of 0-80 mmol/mol. Pressure compensation and temperature regulation of the optical
bench minimize zero and span drift. The optical path can be removed and cleaned by the user if necessary, without
requiring factory recalibration.
Auxiliary Sensor Interface
Adding auxiliary sensors to the LI-8100 is enabled by an Auxiliary Sensor Interface that attaches to the side of the
Analyzer Control Unit. The Interface enclosure is O-ring sealed and has connections for up to 4 thermocouples (types E, J, T, or raw) and 4 general
purpose input voltage channels, any of which can be configured to measure soil
temperature or moisture. Sensors can be powered externally or by the LI-8100
with a constant 5 VDC source. The Auxiliary Sensor Interface also has a 10.5-28 VDC input for use with
external power.
Wireless Communications
The LI-8100 is Wi-Fi (802.11b) enabled, allowing for wireless communication with a compatible handheld
PDA. The LI-8100 Interface Software for PalmŪ OS makes it simple to perform all setup and monitoring functions
via wireless communication or with the included serial cable. LI-COR offers a complete handheld package, which
provides everything you'll need to set up and monitor the LI-8100 with wireless communications, including a PDA.
DC Power
In single chamber configurations, the LI-8100 ACU is powered by either an internal battery (10.5-15 VDC, 3A max.), available
from LI-COR, or a user-supplied DC power source (10.5-28 VDC, 3A max.) connected via the Auxiliary
Sensor Interface. Batteries are rechargeable and provide about 3 hours of
charge, depending on usage. With two battery connectors inside the LI-8100
case, batteries can be changed easily with uninterrupted operation by
connecting a fully charged battery to one connector before removing the
discharged battery from the second connector.
Onboard Data Storage
The LI-8100 has onboard flash storage available for data collection, compressed
at an approximate ratio of 3:1. With compression, about 32 MB of data can
be stored on the instrument. If more space is required for data logging, a
compact flash card can be used as well. LI-COR includes an industrial grade 1 GB compact flash
card, rated for high temperature.