Summer and Winter Settings for LI-7500A and LI-7200 Gas Analyzers

Summer and Winter Settings for LI-7500A and LI-7200 Gas Analyzers

The LI-7200 and LI-7500A gas analyzers regulate the temperature of the chopper motor housing to ensure stable performance over the entire operating temperature range of -25 to +50 °C. The Settings tab of the LI-7200/LI-7500A Windows application software has an Advanced tab that provides options for selecting summer (30 °C) and winter (5 °C) settings for the chopper motor housing temperature.

These settings are not directly related to the temperature requirements for the normal operation of the instrument, but are…

Read More Read More

LI-1800 End of Service

LI-1800 End of Service

LI-COR Biosciences has a legacy of designing and manufacturing robust field instrumentation with long operational lifetimes.  These long operational lifetimes couldn’t be possible without accessible and affordable factory service, skilled technicians and post-sale technical support.  However, it is now time to plan for an end-of-service date of December 31, 2012 for the LI-1800 and LI-1800UW Portable Spectroradiometers, and related accessories.  All service, including recalibration, will be discontinued after December 31, 2012.  Limited replacement parts will be available until our current…