LI-7200 Specifications
- Data Storage: Removable USB storage device (4 GB industrial grade USB flash drive provided, 16 GB available)
- Data Communication: Ethernet, Synchronous Devices for Measurement (SDM; up to 50 Hz), RS-232 (115,200 baud; 20 records per second max), 6 DACs (single ended; 0-5 V)
- Inputs: Ethernet, 4 analog input channels (differential; bi-polar; ±5 V; 300 Hz)
- Operating Temperature Range: -25 to 50° C (-40° C verification test available on request)
- Power Requirements: 10.5 to 30 VDC
- Power Consumption: 12 W nominal (including LI-7550 Analyzer Interface Unit)
- Type: Absolute, non-dispersive infrared gas analyzer
- Detector: Thermoelectrically cooled lead selenide
| CO2 | H2O | |
|---|---|---|
| Calibration Range | 0-3000 µmol mol-1 | 0-60 mmol mol-1 |
| Accuracy | 1% of reading nominal | Within 2% of reading nominal |
| Zero drift (per °C) | ±0.1 ppm typical ±0.3 ppm max. |
±0.03 ppt typical ±0.05 ppt max. |
5 Hz RMS noise:10 Hz 20 Hz (typical @370 µmol mol-1 CO2 |
0.08 µmol mol-1 0.11 µmol mol-1 0.16 µmol mol-1 |
0.0034 mmol mol-1 0.0047 mmol mol-1 0.0067 mmol mol-1 |
| Gain drift (% of reading per °C) |
±0.02% typical ±0.1% max. @ 370 µmol mol-1 |
±0.15% typical ±0.30% max. @ 20 mmol mol-1 |
| Direct sensitivity to H2O (mol CO2/mol H2O) |
±2.00E-05 typical ±4.00E-05 max. |
--- --- |
| Direct sensitivity to CO2 (mol H2O/mol CO2) |
--- --- |
±0.02 typical ±0.05 max. |
- Path Length: 12 cm
- Optical Cell Volume: 16 cm3
- Bandwidth: 5, 10, or 20 Hz, user-selectable
- User Interface: Windows based
- LI-7550 Analyzer Interface Unit (include) Dimensions:
- Size: 35 cm x 30 cm x 15 cm (13.8" x 12" x 6") external dimensions
- Weight: 4.4 kg (10 lbs)
- LI-7200 CO2/H2O Analyzer Dimensions:
- Size: 7.5 cm (3") diameter, 31 cm (12.2") length
- Weight: 1.8 kg (3.95 lbs)
- Cable Length: 5 m (16.4 ft)
- 7200-101 Flow Module (optional):
- Operating Temperature Range: -25 to 50° C
- Power Requirements: 10.5 to 30 VDC
- Power Consumption:
Tube
LengthTube Outer Diameter 0.625 cm
(included)0.953 cm 0.5 m 14 W 10 W 1.0 m 16 W 11 W - Flow Rate: 15 LPM nominal (10 to 18 LPM, user settable)
- Dimensions: 35 cm x 30 cm x 15 cm (13.8" x 12" x 6")
- Weight: 6.15 kg (13.55 lbs)
Specifications subject to change without notice.