LI-7500A Specifications

    CO2H2O
    Calibration range 0-3000 µmol mol-1 0 - 60 mmol mol-1
    Accuracy Within 1% of reading Within 2% of reading
    Zero drift (per °C) ±0.1 ppm typical
    ±0.3 ppm max.
    ±0.03 ppt typical
    ±0.05 ppt max.
    5 Hz
    10 Hz
    20 Hz
    RMS noise: (typical @ 370 ppm CO2 and 10 mmol mol-1 H2O)
    0.08 ppm
    0.11 ppm
    0.16 ppm
    0.0034 ppt
    0.0047 ppt
    0.0067 ppt
    Gain drift
    (% of reading per °C)
    ±0.02% typical
    ±0.1% max.
    @ 370 ppm
    ±0.15% typical
    ±0.30% max.
    @ 20 ppt
    Direct sensitivity to H2O
    (mol CO2/mol H2O)
    ±2.00E-05 typical
    ±4.00E-05 max.
    ---
    ---
    Direct sensitivity to CO2
    (mol H2O/mol CO2)
    ---
    ---
    ±0.02 typical
    ±0.05 max.
  • Data Storage: Removable industrial grade USB flash storage device (4 GB provided, addressable capacity > 16 GB).
  • Data Communication: Ethernet, Synchronous Devices for Measurement (SDM; >50 Hz), RS-232 (115,200 baud; 20 records per second max), 6 DACs (0-5 V; 300 Hz).
  • Inputs: Ethernet, 4 analog input channels.
  • Operating Temperature Range: -25 to 50°C (-40 to 50 °C verification test available on request).
  • Power Requirements: 10.5 to 30 volts DC.
  • Power Consumption: 12 W nominal
    (up to 30 W during startup).
  • Type: Absolute, non-dispersive infrared gas analyzer.
  • Detector Thermoelectrically cooled lead selenide.
  • Bandwidth: 5, 10, or 20 Hz, user-selectable.
  • User Interface: Windows® based.
  • LI-7550 Analyzer Interface Unit Dimensions:
    • Size: 35 cm x 30 cm x 15 cm (13.8” x 12” x 6”) external dimensions.
    • Weight: 4.4 kg (10 lbs).
  • LI-7500A Open Path CO2/H2O Analyzer Dimensions:
    • Size: 6.5 cm (2.6” diameter, 30 cm (12”) length.
    • Weight: 0.75 kg (1.65 lbs.).
    • Cable length: 5 meters (data and IRGA cables).
  • 7550-101 Auxiliary Sensor Interface:
    • Inputs: 4 general purpose ±5 V.
    • Size: 11.5 cm x 6.5 cm x 4.2 cm (4.5” x 2.6” x 1.7”).
    • Weight: 0.39 kg (0.85 lbs) including mounting bracket
  • Specifications subject to change without notice.