LI-610 Portable Dew Point Generator
Used to calibrate H2O analyzers and relative humidity sensors. Includes 610-01 AC Module and one set of 610-04 BNC to mV Recorder Leads. Requires 6200B Rechargeable Battery and LI-6020 Battery Charger for battery powered operation.
[x]Close
Request a Quote
Thank you for your request.
You will receive an email shortly.
If you do not receive it within 24 hours or if you have any questions, please contact LI-COR at 1-402-467-3576.
LI-610 Specifications
- Dew Point Range: 0 to 50 °C (limited to 35 °C below the cooling water temperature)
- Accuracy: ± 0.2 °C (0-50 °C)
- Stability: <0.02 °C per day at 25 °C typical; <0.04 °C per day at 25 °C maximum
- Noise Level: 0.01 °C peak-to-peak
- Repeatability: 0.01 °C
- Response Time: Typically 15 seconds per °C when changing from ambient to a higher dew point; 30 seconds per °C for dew points lower than ambient
- Temperature Sensor: Platinum resistance temperature detector (RTD)
- Flow
- Flow Rate: Adjustable; typically 2.0 liters per minute
- Flow Meter Type: Dwyer series RMA. 2.5 liters per minute full scale
- Flow Meter Accuracy: ± 4% of full scale reading
- Flow Outputs: Two rapid connect hose fittings for 4 mm ID by 6 mm OD plastic tubing. Accepts 1/4" OD tubing with 1/8 to 3/16" ID
- Maximum Input Flow: 2 liters per minute. Contact LI-COR about using higher input flow rates
- Analog Output: 0 to 5 V, 100 mV / °C
- Command Output: 0 to 5 V, 100 mV / °C
- Display: 4 1/2 digit LCD for displaying dew point set temperature (°C), actual dew point temperature (°C), or battery voltage
- Display Resolution: 0.01 °C
- Operating Temperature: 0 to 50 °C, 0 to 100% RH
- Power Requirements: 10.5 to 16 VDC, 5.5 amps max current draw; or 108-126/216-252 VAC with 610-01 AC Module
- Dimensions: 23.5 H x 21 W x 28.5 D cm (9 H x 8.1 W x 11" D)
- Weight: 7.86 kg (17.4 lbs.)