LI-7000 Specifications

CO2 Specifications

  • Range: 0-3000 ppm
  • Accuracy: Within 1% nominally.
  • Zero Drift (with temperature): ±0.3 ppm/°C.
  • Span Drift: ±0.2% of reading/°C.
  • Water Sensitivity: <0.1 ppm CO2/mmol/mol H2O (Software algorithm corrects for band broadening effects).
  • Calibration: Factory linearization over 0-3000 ppm range using NIST-traceable standard gases.
  • Pressure Range: 0 to 115 kPa absolute.
  • Display Resolution: 0.01 µmol/mol.
  • Signal Noise: See Table 1.

H2O Specifications

  • Range: 0-60 mmol/mol. Optical bench and interconnecting hoses must be maintained above the dew point temperature by the user.
  • Accuracy: 1%.
  • Zero Drift: ±0.02 mmol/mol/°C.
  • Span Drift: ±0.4% of reading/°C.
  • Signal Noise: See Table 1.
Noise Specifications (Typical)
 CO2 (370 ppm)H2O (10 mmol/mol)
Average Time (s)Bandwidth (Hz)RMS Noise (ppb)Peak-to-Peak (ppb)RMS Noise (ppm)Peak-to-Peak (ppm)
200.0256390.43
10.525173212
0.5135245217
02015710961175

Table 1. Noise Specifications.