LI-820 Specifications

  • Measurement Range: 0-20,000 ppm
  • Accuracy: <3% of reading
  • Zero Drift (1) : <0.15 ppm/°C
  • Span Drift (2) : <0.03%/°C
  • Total Drift (3) : at 370 ppm: <0.4 ppm/°C
  • RMS Noise at 370 ppm with 1 sec signal filtering: <1 ppm
  • Measurement Principle: Non-Dispersive Infrared
  • Traceability: Traceable gases to WMO standards from 0 to 3,000 ppm. Traceable gases to EPA protocol gases from 3,000 to 20,000 ppm
  • Pressure Compensation Range: 15 kPa - 115 kPa
  • Maximum Gas Flow Rate: 1 liter/min
  • Output Signals: Two Analog Voltage (0-2.5V or 0-5V) and Two Current (4-20mA) Digital: TTL (0-5V) or Open Collector
  • DAC Resolution: 14-bits across user-specified range
  • Source Life: 18,000 Hours (~2 years continuous use)
  • Power Requirements: Input Voltage 12-30 VDC; 1.2A @ 12V (14W) maximum during warmup with heaters on; 0.3A @ 12V (3.6W) average after warmup with heaters on
  • Operating Temperature Range: -20 to 45°C
  • Relative Humidity Range: 0 to 95% RH, Non-Condensing
  • Dimensions: 8.75" x 6" x 3" (22.23 x 15.25 x 7.62 cm)
  • Weight: 2.2 lbs. (1 kg)

(1) Zero Drift is the change with temperature at 0 concentration.
(2) Span Drift is the residual error after re-zeroing following a temperature change.
(3) Total Drift is the change with temperature without re-zeroing or re-spanning.