LI-840A Specifications
CO2
- Measurement Range: 0-20,000 ppm
- Accuracy: < 1% of reading
- Calibration Drift
- Zero Drift1: <0.15 ppm/°C
- Span Drift2: <0.03%/°C
- Total Drift3 at 370 ppm: <0.4 ppm/°C
- RMS Noise @ 370 ppm with 1 second signal filtering: <1ppm
- Sensitivity to water vapor: <1.0 ppm CO2/mmol/mol H2O
H2O
- Measurement Range: 0-60 mmol/mol
- Accuracy: < 1.5% of reading
- Calibration drift
- Zero Drift1 at 0 mmol/mol: <0.003 mmol/mol°C
- Span Drift2 at 10 mmol/mol: <0.03%/°C
- Total Drift3 at 10 mmol/mol: <0.009 mmol/mol/°C
- RMS Noise at 10 mmol/mol with 1 sec signal filtering: <0.01 mmol/mol
- Sensitivity to CO2: <0.0001 mmol/mol H2O/ppm CO2
- Measurement principle: Non-Dispersive Infrared
- Traceability: Traceable to gases to WMO standards for CO2. NIST traceable LI-610 Portable Dew Point Generator for H2O
- Pressure Compensation Range: 15 kPa-115 kPa
- Maximum Gas Flow Rate: 1 liter/min
- Output Signals: Two analog voltage (0-2.5 V or 0-5 V) and two current (4-20mA)
- Digital: TTL (0-5V) or Open Collector
- DAC Resolution: 14-bit across user-specified range
- Source Life: ˜18,000 Hours
- Power Requirements: Input Voltage 12-30 VDC
- 1.2 A @ 12V (14W) maximum during warm-up with heaters on
- 0.3A @ 12V (3.6W) average after warm-up with heaters on
- Operating Temperature Range: -20°C to +45°C
- Relative Humidity Range: 0 to 95% RH, Non-condensing
- Dimensions: 8.75" x 6" x 3" (22.23 x 15.25 x 7.62 cm)
- Weight: 2.2 lbs (1 kg)
- Internal Optical Cell Volume: 14.5 mL
1 Zero drift is the change with temperature at 0 concentration
2 Span drift is the residual error after re-zeroing following a temperature change
3 Total drift is the change with temperature without re-zeroing or re-spanning
*Specifications subject to change without notice