COVID-19 Update: We remain open and ready to help. Read More

Specifications

Laser/Microscope

  • Image Field Size
    10 cm × 12 cm
  • CCD Pixel Size
    6.45 microns
  • Dynamic Range*
    22 bit (>6 logs)
  • Depth of Field for Best Sample Focus
    6 mm
  • Patented FieldBrite XT2 Technology
    CV <3% across field
  • Laser Lifetime
    20,000 hours of operation
  • 700 Channel Laser Source
    Solid-state diode laser at 685 nm
  • 800 Channel Laser Source
    Solid-state diode laser at 785 nm
  • Light Source
    600 nm diffused light source
  • Detectors
    Low-noise CCD. Thermoelectrically cooled.
  • Acquisition Times
    Fluorescence (700 and 800 nm) channels: 30 s, 2 min, 10 min plus variable time feature. Chemiluminescence channel: 30 s, 2 min, 10 min, 60 min plus variable time feature.
  • Focusing
    Automatic

* This dynamic range is obtained in a single acquisition.

Operation Specifications

  • Operating Conditions
    For indoor use only; operating temperature 15-35 °C and dew point no greater than 22 °C, non-condensing; maximum operating temperature may be reduced at elevations above 2000 m.
  • Power Requirements
    Universal input range is between 100-127 VAC and 200-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); 4 Amp maximum; at 120V, 0.3 Amp typical when idle; 50/60 Hz. Overvoltage Category II.

Size and Weight

  • Dimensions (instrument only)
    41.4 cm W × 47 cm D × 67.3 cm H
    (16.3" × 18.5" × 26.5")
    Depth with imaging drawer open is 59.7 cm (23.5")
  • Weight
    27 kg (60 lbs)

Software Specifications

  • Empiria Studio® Software
    Requires a minimum of 8 GB of RAM; 64-bit versions of Windows 7 or 10; Mac OS Sierra or High Sierra
  • Image Studio Software
    For operating system compatibility, see this table.
  • Please contact LI-COR for compatibility with other software packages.

Ask us about our solutions for your regulated lab including IQ/OQ and 21 CFR part 11-ready software.

chat iconChat With Us
chat icon disabled