Chat With Us

Specifications

Anesthesia Suite

SmartFlow Vaporizer

  • Oxygen Flush Volume:
    30-35 L/m
  • Circuit Delivery Volume
    • Induction Chamber:
      1 L/m ± 10%
    • Breathers 1-3 (Pearl® Impulse, Surgical Suite, Docking Station):
      0.5 L/M ± 10%
    • Breather 4 (MousePOD):
      1.5 L/m ± 10%
  • Gas Pressure Requirements:
    45-55 psi (50psi optimum)
  • System Pressure Gauge Range:
    0-15 psi
  • Dimensions (WxHxD, inches):
    13.5 x 11 x 6
  • Weight:
    ~23 lb.
  • Material:
    • Stand:
      Galvanized and Annealed Steel
    • Vaporizer:
      Brass/Aluminum
  • Gas Compatibility:
    Isoflurane
  • Recommended service intervals:
    12-24 months

IMPORTANT NOTE: The Vaporizer on the SmartFlow system is custom designed and MUST NOT be used with any other anesthesia system.


SmartFlow Induction Chamber

  • Material:
    Acrylic (Do NOT clean with alcohol)

SmartFlow Water Pump

  • Dimensions (inches):
    11.5 x 8 x 8
  • Weight (empty):
    6.5 lb
  • Reservoir capacity:
    93 ox (2.75L)
  • Flow Rate:
    9 gph (34 lph)
  • Ambient operating temp:
    60°F - 90°F
  • Temperature Setpoints:
    107°F (42°C)
    100°F (38°C)
    95°F (35°C)
    50°F (10°C)
  • Average Temperature Accuracy:
    ±2°F at 107°F (±1°C at 42°C)
  • Maximum Contact Temperature:
    107°F (42°C)
  • Electrical Requirements (120V)
    (For SmartFlow Model 9000-101):
    • Voltage (VAC):
      120±10%
    • Frequency (Hz):
      60
    • Current (Amps):
      3.1 amperes
  • Electrical Requirements (230V)
    (For SmartFlow Model 9000-102):
    • Voltage (VAC):
      230±10%
    • Frequency (Hz):
      50
    • Current (Amps):
      2.0 amperes

Surgical Suite

SmartFlow Surgical Bed

  • Dimensions:
    10.0"L x 3.7"W x 1.0"H
  • Material:
    Heavy gauge stainless steel

SmartFlow Microflex Rebreather

  • Animal size range:
    Up to 500 g

Scroll to Top