Specifications

Multiplexer

LI-8150 Multiplexer

  • Dimensions: 40.6 cm L × 57.2 cm W × 21.1 cm H (16” × 22.5” × 8.3”)
  • Weatherproof Rating: Tested to IEC IP55 standard
  • Weight: 9.4 kg (20.7 lbs), 8 ports; 11.2 kg (24.8 lbs) 16 ports
  • Operating Temperature Range: -20 to 45 °C
  • Operating Humidity Range: 0 to 95% RH, non-condensing
  • Maximum Spread: 30.0 m (98.4 ft)
  • Flow Rate to Chambers: ~3.0 lpm (non-adjustable)
  • Flow Rate Between LI-8100A and LI-8150: ~2.0 lpm
  • Power Requirements: 10.5 VDC to 14.5 VDC (120 VAC and 240 VAC with optional power supply)
  • Soil Temperature Thermistor (optional): ±1.0 °C from -20 to 50 °C

See table below for total system power requirements

LI-8100 + LI-8150 + (n) Chambers Sampling: No Chamber Movement Sampling: Two chambers move at one time Warm-up: Up to four chambers move at one time
(n) Amps @ 12.5VDC Watts Amps @ 12.5VDC* Watts Amps @ 12.5 VDC* Watts
1 1.0 12.5 3.0 37.5 3.0 37.5
2 2.0 25.0 3.8 47.5 3.8 47.5
4 2.1 26.3 3.8 47.5 4.8 60
8 2.1 26.3 3.8 47.5 4.8 60
16 2.3 28.8 3.8 47.5 4.8 60

*Power requirements of the LI-8100/8150 System at minimum operating temperature.


8150-770 AC to DC Power Supply

  • Power Requirements: 115-120VAC or 230-240VAC, 50/60 Hz, 230W. Input voltage range is switch-settable.
  • Weatherproof Rating: Tested to IEC IP55 standard
  • Output Voltage: 12 VDC
  • Operating Temperature Range: -20 to 50 °C (-4 to 122 °F)
  • Dimensions: 18.1 cm D × 28 cm L × 10.2 cm H (7.125" × 11" × 4")
  • Weight: 3.1 kg (14.05 lbs)