LI-3050C Specifications

Resolution: 1 mm2 (1 mm x 1 mm scanning area).

Accuracy: Within ± 1% for samples > 10 cm2.

Sample Dimensions

Width: 127 mm maximum, 1 mm minimum.

Thickness: 7.5 mm maximum.

Length: unlimited.

Conveyer Belt Speed: 6.3 cm/s at 60 Hz; 5.4 cm/s at 50 Hz.

Transparent Belts: Rugged clear vinyl.

Power Requirement: 20 Watts maximum, 115/230 VAC, 50/60 Hz.

User-Replaceable Mains Fuse: 2/10 amp type T for 115 VAC, 1/10 amp type T for 230 VAC, ¼ x 1 ¼ glass cartridge style.

Operating Temperature: 0 to 55 °C. <95% RH (non-condensing).

Storage Temperature: -20 to 55 °C. <95% RH (non-condensing).

Size: 27.9 cm H × 27.9 W × 38.1 L (11.0" × 11.0" × 15.0")

Weight: 7.7 kg (17.0 lb).