Specifications

Resolution: 1 mm2 or 0.1 mm2 (user-selectable).

Scanning Area:

1 mm2 Resolution: 1 mm × 1 mm.

0.1 mm2 Resolution: 0.300 mm L × 0.33 mm W.

Accuracy:

  Sample Area
Resolution 10 cm2 5 cm2 1 cm2 0.5 cm2 0.25 cm2
1 mm2 ±1% ±2% ±5% ±7% --
0.1 mm2 ±0.5% ±1% ±1% ±1.5% ±4%

Combined accuracy and precision to 99% confidence with correct calibration on verifiable shapes. Accuracy can be increased by calibrating and using in a specific zone of measurement (e.g. the center). Samples with a high proportion of edges, such as conifers, roots, wires, etc. will raise uncertainties to 5% and more. Use only 0.1 mm2 resolution for these types of samples.

Display Capacity:

1 mm2 Resolution: 999,999.99 cm2

0.1 mm2 Resolution: 99,999.999 cm2

Display: 8-digit LED (light emitting diode).

Sample Dimensions:

Width: 25.4 cm maximum; 1.5 to 3 mm minimum.

Thickness: Up to 2 cm, expandable to 2.5 cm by the user.

Length: Unlimited.

Conveyer Belt Speed: 8.0 cm s-1 at 60 Hz; 6.7 cm s-1 at 50 Hz.

Transparent Belts: Rugged clear vinyl.

Light Source: 15 watt fluorescent tube.

Power Requirement: 108-126/216-252 VAC, 48-66 Hz, 100 W max.

Operating Temperature: +15 to 55 °C.

Storage Temperature: -20 to 65 °C.

Size: 25.0 H × 60.0 W × 73.0 cm L (9.8" × 23.6" × 28.7")

Weight: 43 kg (95 lb.).