Specifications
Specifications are subject to change without notice.
Power Supply Voltage: 10 - 54 VDC
Power: 400 mW when no sensor is connected; up to 12 W maximum
Environment: Suitable for outdoor use in wet locations
Altitude: Up to 5000 meters
Temperature: -20 to 50 °C
Relative Humidity: 0 to 100%.
Transient Overvoltage: Category II
Temporary Overvoltage: 60 V
Pollution Tolerance: Level 3 (enclosure must be intact and in good condition)
Four Serial Ports for TriSonica Anemometers:
UART-TTL or EIA232 voltage level
Configurable baud rate and parity
12V output to power TriSonica sensors
Trigger output to synchronize multiple sensors
Instrument Ports:
LI-550P and LI-560 Trigger Output
RS-232
GPS Port and External Radio Port:
NMEA0183 GPS
TTL-UART
I2C
JST 6-pin GH connector
GPS can be timing standard
MicroSD Card Storage:
Formats: FAT16, FAT32, exFAT
Maximum Card Capacity: >32 GB
Data Sampling:
Maximum scan rate: 100 Hz
Adjustable sampling rates
User-definable response for missing data
Real-Time Clock with Time Tagging Option
External Connections ESD Protected
Weight: 150 g (5.3 oz.) early models weighed up to 450 g (16 oz.)
Dimensions: