Configuring the CSAT3 to Log EC Datasets in the LI-7200/RS
Printable PDF: Configuring the CSAT3 to Log EC Datasets in the LI-7200/RS
Instructions for configuring the CSAT3 for LI-COR EC systems.
The CSAT3 sonic anemometer (Campbell Scientific, Inc., Logan, UT) can be configured to log raw flux data in conjunction with an LI-7500RS or LI-7200RS gas analyzer from LI-COR in a compressed .ghg file format. The .ghg file format makes it possible to continuously collect raw flux data for more than two months on a single 4 GB USB drive and to process data conveniently with LI-COR’s EddyPro flux data processing software.
- The CSAT3 is connected to an LI-7550 Analyzer Interface Unit through analog output/input channels. There are three cabling options for interfacing the CSAT3 to the LI-7550:
- The LI-7550 to CSAT3 interface cable (p/n 9975-035) provides a direct cable connection (shown below). This cable is not included with the LI-7500RS and LI-7200RS but can be ordered separately.

- The 9975-035 cable from LI-COR can be used to connect the CSAT3 to the LI-7550.
- The CSAT3 analog output cable can be wired into the LI-7550 auxiliary sensor interface box (LI-COR p/n 7550-101).
- The LI-7550 analog input/output cable (p/n 392-10109) can be spliced into the CSAT3 analog output cable, or soldered into the CSAT3’s analog connector.
- Analog outputs from the CSAT3 are turned off by factory default. To turn them on, use the CSAT32 software, which can be downloaded from:
- http://www.campbellsci.com/19_1_9999_68
- Set the Measurement Rate of the CSAT3 to 60 Hz for better data synchronization with the outputs from the gas analyzers.
- Scale the CSAT3 analog outputs. Choose Low or High and set the offsets and multipliers for the output range.
-
Slope, Low range (m s-1 v-1) Slope, High range (m s-1 v-1) Offset (m s-1) ux 6.5536 13.1072 0 uy 6.5536 13.1072 0 uz 1.6384 1.6384 0 c 6.5536 13.1072 340 - Follow the quick start guide to set up LI-COR gas analyzers to log eddy covariance data in the .ghg file format. The CSAT3 provides the analog outputs as U, V, W, and Speed of Sound (SOS). The LI-7550 will need to be configured to read SOS, rather than sonic temperature (Ts), for a CSAT3.