Specifications

General

  • Weight: 2.02 kg
  • Power Requirements
    • Voltage: 9 to 33 VDC nominally; Can be powered from a standard USB port (2.0 or newer) for calibration and firmware updates. Outdoor operation from a 5 V USB power supply is not supported.
    • Power Consumption: 1.5 W
  • Communication Interfaces: SDI-12, RS-232 (Raw 10Hz data)
  • Other Devices:
    • Integrated GPS
    • Positional Accuracy: Provides location and precision time, including for edge flux processing.
    • Supports GPS Time Output: GPST
    • Magnetometer (Output instrument orientation relative to magnetic north)
    • Accelerometer (Output instrument tilt)
  • Embedded Software: Capable of software updates via USB
  • Operating Temperature Range: -25 - 50 °C

CO2 Measurement

  • Calibration Range: 0 to 1500 µmol mol-1
  • Accuracy: Within 1.5% of reading
  • Zero Drift (per °C): ±0.15 ppm typical, ±0.3 ppm maximum
  • RMS Noise: @10 Hz: 1.0 ppm (typical @ 400 µmol mol-1 CO2)
  • Direct Sensitivity to H2O: ±2.00E-05 typical, ±4.00E-05 maximum

H2O Measurement

  • Calibration Range: 0 to 60 mmol mol-1
  • Accuracy: Within 1.5% of reading
  • Zero Drift (per °C): ±0.03 mmol mol-1 typical, ±0.05 mmol mol-1 maximum
  • RMS Noise: @10 Hz: 0.05 mmol mol-1 (typical @ 10 mmol mol-1 H2O)
  • Direct Sensitivity to CO2: ±0.02 typical, ±0.05 maximum

Wind Speed

  • Measurement Axes: U, V, W
  • Measurement Range: 0 - 30 m s-1 (Horizontal Wind Conditions)
  • W Offset at Zero Wind: ±0.06 m s-1
  • W RMS Noise: <0.1 m s-1 @ 5 m s-1, <0.15 m s-1 @ 15 m s-1
  • Sonic Temperature Accuracy: ±0.7°C Maximum offset at 20°C

Biomet Measurements

  • Photosynthetic Photon Flux Density:
    • Range: 0-3000 µmol m-2 s-1
    • Accuracy: ±5% of reading
    • Cosine Correction: Corrected up to 75° angle of incidence
  • Air Temperature:
    • Range: -40 to 60°C
    • Accuracy: ±1.5°C - No load conditions
  • Atmospheric Pressure:
    • Range: 50 - 110 kPa
    • Accuracy: ±0.2 kPa
  • RH:
    • Range: 0-100% - non-condensing
    • Accuracy: ±1% typical

Cellular

  • LTE CAT1
  • Data Transfer Frequency: Data sent every half-hour
  • Regional Approvals:
    • USA
    • EU
    • UK
    • Norway
    • Greenland
    • Switzerland
    • Canada
    • Japan
    • Argentina
    • Kazakhstan
    • Mexico
    • Morocco
    • Philippines

Solar Power

  • Solar Panel: 100 Watt Monocrystalline
  • Battery: 12V, 100Ah AGM
  • Battery Weight: 25Kg (55 lbs.)
  • Charger: 9A MPPT (Maximum Power Point Tracking) Charger

Auxiliary Power Input

  • Input Voltage Range: 9 to 33 V (>11 VDC recommended for startup)
  • Maximum Current: 3 A

Digital Inputs/Outputs

  • SDI-12 Eddy Covariance Sensor: 1
  • SDI-12 ports for other sensors: 1
  • Output Voltage: 12V
  • Maximum Total Current: 2.5A
  • Micro USB port for firmware updates
  • HOBOnet (Wireless Biomet) port

Mast

  • Maximum Height w/ LI-720: 5M (198 in.)
  • Maximum Load on Mast 4.54 kg (10 lbs.)

Data

  • Data Storage: Removable Industrial Micro SD card; 8 GB.
  • Message Buffer: If the modem is offline, data is stored until connectivity is restored.
  • Data Format: Text, Comma-separated Values (CSV)
  • GNSS (GPS) Support: GPS receiver for time synchronization and location information

Specifications subject to change without notice.