This document describes mounting hardware that is used to install a sonic anemometer and the LI-7500A/RS/DS, LI-7200/RS, and LI-7700 gas analyzers LI-COR eddy covariance system. A typical installation is to a 3.3 cm (1.315 inch) outside diameter (OD) cross arm. Dimensions given refer to Iron Pipe Size (IPS) standards and will differ from the measured dimensions.
| Description | Part Number | |
|---|---|---|
| Crossover Fitting; ¾" × 1" IPS; Pipe OD: 26 mm (1.05") × 33 mm (1.315") |
|
7900-342 |
| Crossover Fitting; 1" × 1" IPS; Pipe OD: 26 mm (1.05") × 33 mm (1.315") |
|
7900-346 |
| Swivel Mount; ¾" IPS; Pipe OD: 26 mm (1.05"). |
|
7900-344 |
| 56 cm Riser Bar; ¾" IPS, 56 cm (22") Long; OD: 26 mm (1.05") | 9879-046 | |
| 45 cm Riser Bar; 1" IPS, 45 cm (18") Long; OD: 33 mm (1.315") |
9879-095 |
Set screws should be tightened to 22 N•m (16 ft•lbs). If no torque wrench is available, tighten each set screw until it contacts the pipe, then one more full revolution for aluminum pipe, or ¼ revolution for stainless steel. Use a 5/32" hex key for all the fittings except the ¾" × 1" crossover fitting (7900-342), which uses a 3/16" hex key.
Mounting kits
Mounting hardware kits are available for gas analyzers and a variety of sonic anemometers.
Gas analyzer mounting kit
| Description | Quantity | Part Number |
|---|---|---|
| Gas Analyzer Mounting Kit | - | 7900-340 |
| Swivel Mount (¾") | 1 | 7900-344 |
| 56 cm Riser Bar (¾" IPS) | 1 | 9879-046 |
| Crossover Fitting (¾" × 1") | 1 | 7900-342 |
The combination mount (7900-320) is for mounting a Gill WindMaster/Pro and LI-7500A/RS/DS gas analyzer to a tripod cross-arm (requires 900-11242 and 7900-342). The gas analyzer mounting kit (7900-340) includes alternative hardware for installing components on a tripod cross-arm. For the LI-7200/RS, use the mounting kit plus one 56 cm riser bar (9879-046) and crossover fitting (7900-346) to support the intake tube.
Gill WindMaster/Pro anemometer combination mount
The combination mount (7900-320) can be used with the anemometer mount (900-11242) and crossover fitting (7900-320).
Gill WindMaster/Pro anemometer mounting kit
To mount the WindMaster/Pro and R3, use a ¾" × 1" cross-over fitting (7900-342) and anemometer mounting post (900-11242). The mounting post is included with WindMaster/Pro anemometers purchased from LI-COR. The cross-over fitting is available seperately.
| Description | Quantity | Part Number |
|---|---|---|
| WindMaster/Pro Mounting Post | 1 | 900-11242 |
| Crossover Fitting (¾" × 1") | 1 | 7900-342 |
RM Young 81000x anemometer mounting kit
Mount the RM Young 81000, 81000V, 81000RE, or 81000VRE sonic anemometer with the RM Young mounting kit (7900-332).
| Description | Quantity | Part Number |
|---|---|---|
| Crossover Fitting (1" × 1") | 1 | 7900-346 |
| 45 cm Riser Bar (1" IPS) | 1 | 9879-095 |
Metek MultiPath Class A anemometer mounting kit
Mount the Metek Class A anemometer with the Metek mounting kit (7900-336).
| Description | Quantity | Part Number |
|---|---|---|
| Crossover Fitting (1" × 1") | 1 | 259-17893 |
| 45 cm Riser Bar (1" IPS) | 1 | 9879-095 |
| Cross-arm Plate | 1 | 9879-020 |
| Cross-arm Brackets | 2 | 9879-043 |
| Cross-arm Hardware Pack | 1 | 9979-018 |
Metek MultiPath Cage anemometer mounting kit
Mount the Metek Cage anemometer with one 1" × 1" crossover fitting (7900-346).