Factory Calibration
LI-7500DS Recommended Recalibration Frequency: Every two years; however, if the internal chemicals are serviced every two years and the analyzer zero and span are verified, then no regular factory recalibration is required.
What's included with factory recalibration?
A factory recalibration includes the following:
Calibration
- Analyzer drift and stability is checked.
- H2O analyzer calibration (6 points, 3 temperatures).
- CO2 analyzer calibration (13 gas concentrations at 3 temperatures).
- The accuracy of the calibration function is then verified across the measurement range of CO2 and H2O.
- Pressure transducer calibration.
- Temperature sensors verification.
Tests and Maintenance
- Clean analyzer, if needed.
- Change detector housing chemicals.
- Upgrade software to the most recent version.
- Other repairs and maintenance as needed; you'll be notified if additional repairs are required.
To request factory service or recalibration for an instrument:
Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.
 
                                            


