General Description
What's What
- RS-232 Cable - (part number 392-04993), used to connect the LI-820 to your computer. This cable is a null modem (cross-over) type cable; a straight through type cable will not work with the LI-820. Appendix B shows the pin assignments for the DB-9 connector on the LI-820 front panel.
- 820-500 Data Communications Software - This Windows-compatible software is used to communicate with your computer. The software is provided on CD. A complete description of this software can be found in Section 3, Operation.
- Standard Spare Parts Kit (part number 9980-012) - This kit contains replacement parts for your LI-820. As you become familiar with the analyzer you will learn which items to keep close at hand and which items can be stored away. Some of the items in the spare parts kit include:
| Description | Qty. | LI-COR Part No. |
|---|---|---|
| Spares Kit, includes: | ||
| Polyurethane Mounting Feet | 4 | 234-02268 |
| Mounting Screws | 4 | 149-02610 |
| Terminal Block | 1 | 331-05273 |
| Bev-a-line Tubing | 12' | 222-01824 |
| Balston Air Filter | 2 | 300-01961 |
| Quick Connect Unions | 2 | 300-03123 |
| Rapid Tube Fitting Nut | 2 | 9861-036 |
| 2A Fuse | 2 | 439-03952 |
| Cleaning Kit, includes: | 1 | 9980-013 |
| 5" Cleaning Swabs | 5 | 610-05315 |
| O-Rings | 4 | 192-00226 |
| Pre-formed Optical Bench Tubing | 4 | 6580-010 |
Features
The LI-820 is an economical high performance, non-dispersive infrared gas analyzer designed to be used for a wide variety of applications. Some of the LI-820's important features include:
- The 14 cm (5.5") optical bench is used to measure CO2 concentrations over a range of 0 - 20000 ppm.
- The optical bench is fully serviceable by the user; cleaning the optical bench is a simple process, and does not affect the instrument calibration.
- Simple Windows® software provides for easy user calibration and data output.
Precautions
- Avoid storing the LI-820 in very humid conditions for more than several days. Doing so can cause the optical sensor to become saturated with humidity, which will temporarily affect measurements when the instrument is removed from storage. See for more details.
- The optical bench is maintained at a constant 50 °C. Avoid direct exposure to sunlight or extremely high temperatures that may elevate the temperature inside the LI-820 case.
- The optical source is sensitive to vibration, and can be damaged by strong vibrations or jarring. Do not drop the LI-820, or expose it to severe mechanical shock.
- Do not use abrasive cleansers when cleaning the optical bench, as damage to the gold plating may result. Instructions for cleaning the optical bench can be found in Section 5.