Factory Calibration
Recommended Recalibration Frequency: No regular factory recalibration required.
What's included with LI-840A recalibration?
A factory recalibration includes:
Calibration
- Analyzer drift and stability is checked.
- H2O analyzer calibration (6 points, 3 temperatures).
- CO2 analyzer calibration (17 gas concentrations at 3 temperatures).
- The accuracy of the calibration function is then verified across the measurement range of CO2 and H2O.
Tests and Maintenance
- The optical path components and hoses are inspected.
- Software is updated as needed.
- Other repairs and maintenance as needed; customers are notified if additional fees are required for repairs.
To request factory service or recalibration for an instrument:
Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.
 
                                            


