LI-8810 AtmospherIQ™ CH4/CO2/H2O Gas Analyzer

PROVEN PERFORMANCE FOR ATMOSPHERIC BASELINE METHANE MEASUREMENTS

Request Quote
Slanted Shape

Dependable gas analyzers are critical to long-term atmospheric methane measurements. The LI-8810 brings dependable performance and unmatched support to long-term CH4 observatories. Built using the proven OF-CEAS platform and backed by a 5-year no-questions-asked warranty, the LI-8810 is designed to make atmospheric CH4 monitoring more widespread and dependable than ever.

ATMOSPHERIC GAS MEASUREMENTS WITH PEACE OF MIND

The optical bench that was tested by leading networks and proven in LI-COR Trace Gas Analyzers is now available for long-term monitoring networks. Performance evaluations show that methane measurements by the LI-8810 meet or exceed requirements for leading networks.

CH4 Analysis SpecificationsWMO Compatibility GoalICOS Station SpecificationsLI-8810 Performance
Precision, 1 σShort Period±2 ppb< 1 ppb (1 min)< 0.25 ppb (5 sec)
Long Period< 0.50 ppb (1 hour)< 0.05 ppb (1 hour)
Repeatability (1 σ), 10 min avg< 0.50 ppb< 0.40 ppb (8 hour)
Table 1. Performance evaluations show that the LI-8810 CH4 measurements meet or exceed WMO goals and ICOS standards for atmospheric methane gas analysis.
AtmospherIQ LI-8810 CH4 Allan Plot
Figure 1. CH4 stability data from an LI-8810 as it measured continuously flowing tank gas (2000 ppb CH4) over a 48-hour period. The Allan deviation plot is the CH4 precision with 1-second averaging. Error bars represent 68% (±1σ) confidence intervals based on the number of averaged time series available at each τ.

LASER-FOCUSED ON CONTEMPORARY MEASUREMENT NEEDS

AtmospherIQ Gas Analyzers feature numerous patented technologies and innovations that ensure high-quality data and dependable long-term operation. The combination of a high-finesse optical cavity, dense data grid created by fixed cavity resonance modes, and tightly controlled cavity temperature and pressure ensures that AtmospherIQ™ Gas Analyzers satisfy the expectations for contemporary laser-based gas analyzers.

5-Year

Protection

Market-leading 5-year service agreement with loaner instruments included, if needed, comes standard with each analyzer. No questions asked.

Patented

Innovation

Patented technologies optimize the proven Optical Feedback – Cavity Enhanced Absorption Spectroscopy (OF-CEAS) platform for atmospheric monitoring.

Versatile

Pump Options

Vibration-isolated internal pump or optional external pump for versatility.

Battery

Backup

Backup battery power ensures continuous logging and elegant shutdown if main power is lost.

5-Year

Protection

Market-leading 5-year service agreement with loaner instruments included, if needed, comes standard with each analyzer. No questions asked.

Patented

Innovation

Patented technologies optimize the proven Optical Feedback – Cavity Enhanced Absorption Spectroscopy (OF-CEAS) platform for atmospheric monitoring.

Versatile

Pump Options

Vibration-isolated internal pump or optional external pump for versatility.

Battery

Backup

Backup battery power ensures continuous logging and elegant shutdown if main power is lost.

5-Year

Protection

Market-leading 5-year service agreement with loaner instruments included, if needed, comes standard with each analyzer. No questions asked.

Patented

Innovation

Patented technologies optimize the proven Optical Feedback – Cavity Enhanced Absorption Spectroscopy (OF-CEAS) platform for atmospheric monitoring.

Versatile

Pump Options

Vibration-isolated internal pump or optional external pump for versatility.

Battery

Backup

Backup battery power ensures continuous logging and elegant shutdown if main power is lost.

PLUG IN

A SIMPLE ADDITION TO ATMOSPHERIC MONITORING STATIONS

AtmospherIQ Gas Analyzers are designed for compatibility with existing monitoring stations.

  • Versatile data management and controls support communication with devices and data management infrastructure.
  • Support for custom software integrations to simplify remote data acquisition, instrument calibration, and system monitoring.
  • Built-in data storage automatically records all measurements, events, and performance excursions for robust data quality analysis and redundancy.
  • Low power operation dispels concerns about marginal main power supplies, with compatibility with global main power standards.
  • Backup battery power ensures continuous logging and elegant shutdown if main power is lost.
  • Simple tubing connections for easy addition to measurement infrastructure.
  • Analyzers mount in a standard 19″ instrument rack, occupying four rack units (4U) each.

LI-8810 Specifications

CH4 Measurements

  • Precision (1σ):
    • 0.25 ppb with 5-second averaging
    • 0.10 ppb with 5-minute averaging
    • 0.05 ppb with 1-hour averaging
  • Drift (1σ), 10 min avg, 8-hr: < 0.4 ppb
  • Response Time (T10-T90) from 0 to 2 ppm:
    • ≤ 2 seconds in Standard Configuration
    • ≤ 3 seconds in High Altitude Configuration
  • Range: 0 to 100 ppm

H2O Measurements

  • Range: 0 to 60,000 ppm
  • Precision (1σ):
    • 20 ppm at 10,000 ppm with 5-second averaging
    • 5 ppm at 10,000 ppm with 5-minute averaging

General

  • Measurement Technique: OF-CEAS (Optical Feedback – Cavity Enhanced Absorption Spectroscopy) direct molecule measurements
  • Measurement Rate: 1 sample per second (1 Hz)
  • Optical Cavity Volume: 6.4 cm3
  • Flow Rate:
    • 250 sccm nominal in Standard Configuration
    • 150 sccm nominal in High Altitude Configuration
  • Total Weight: 18.2 kg (including batteries)
  • Case Dimensions: 56 cm × 48.25 cm × 18 cm (D × W × H)
  • Operating Temperature Range: -25 °C to 45 °C
  • Operating Humidity Range: 0 to 85% RH (non-condensing, without solar load, under normal operating conditions)
  • Sample Line Humidity Range: 0 to 99% non-condensing
  • Operating Pressure Range:
    • 70 to 110 kPa in standard and reduced flow rate configuration
    • 50 to 110 kPa in high altitude configuration
  • Connectivity: Ethernet TCP/IP; RJ45 connector
  • Power Consumption:
    • Steady State Operation: 22 Watts at 25 °C without batteries charging
    • Warmup: Up to 100 W with batteries charging
  • Power Supply Adapter (included): Universal Power Adapter (Input: 100 to 240 VAC, 50-60 Hz; Output: 24 VDC)
  • Battery Life: 8 hours typical with 2 batteries
  • U.S. Patents: 8,659,758; 8,665,442; 8,885,167; 9,116,047; 9,194,742; 9,304,080; 9,581,492; 9,678,003; 9,759,654; 9,759,655; 9,989,729; 10,514,299; 10,527,492 and 11,441,944.

External pump

  • Dimensions: 17 cm × 27 cm × 18 cm (D × W × H)
  • Weight: 4.1kg

Specifications subject to change without notice.