Pearl Trilogy Specifications

Specifications

Operating Conditions: For indoor use only; operating temperature 15-35 °C and dew point < 22 °C, non-condensing; maximum operating temperature may be reduced at elevations above 2000m.

Environmental Conditions: Pollution Degree 2.

Detector Type: CCD. Thermoelectrically cooled.

Acquisition Time: 500 ms to 40 s.

Acquisition Time: 500 ms to 60 s.

Wavelength Maxima for Excitation/Emission Bands:

ChannelExcitationEmission
700 Channel685 nm720 nm
800 Channel785 nm820 nm
White ChannelWhiteNA
BioluminescenceNA450 - 650 nm

Dyes: Optimized for LI‑COR® IRDye® 680RD and IRDye 800CW.

Resolution: 85, 170, and 255 micron.

Image Display Options: Pseudocolor, grayscale, single color (red, green, or blue), or two colors with overlapping fluorescence displayed in a third color.

Imaging Drawer: 16.8 cm W × 12 cm D (6.6” W × 4.75” D). Vertical clearance to top of drawer is 7.5 cm (3”).

Field of View: 11.2 cm W × 8.4 cm D (4.4” W × 3.3” D) at the surface of the imaging stagebed.

Power Requirements: Universal input range is between 100-127 VAC and 200-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); 50/60 Hz; at 120V, 4 Amp maximum; at 120V, 0.3 Amp typical when idle. Overvoltage Category II.

Computer Connection Cable: Cat. 5e RJ45, 10BASE-T/100BASE-TX/1000BASE-T. Use only the supplied cable.

Size: 41 W × 41 D × 66 cm H (16” W × 16” D × 26” H). Depth with imaging drawer open is 63.5 cm (25”).

Weight: 25 kg (55 lb).

The Pearl Imager is certified as a Class 1 laser product. Hazardous laser radiation is not emitted outside the instrument. Radiation emitted inside the Pearl Imager is confined within protective housings and external covers. A series of safety interlocks ensures that laser beams cannot escape during any phase of user operation.

Embedded Firmware Source Code

The embedded firmware running on the Odyssey CLx ImagerOdyssey DLx ImagerPearl ImagerOdyssey Sa ImagerOdyssey Fc ImagerOdyssey XF ImagerOdyssey M ImagerC‑DiGit Blot Scanner (model number 91409142943092602800280235003600) utilizes open source software. To obtain source code or related information, contact LI‑COR Support at http://www.licor.com/biotechsupport.